SelfCalibration
This product is equipped with a built-in calibration sensor. By setting the calibration targets and execution schedule in advance, the calibration sensor automatically operates and regularly calibrates the monitor. This automatic calibration function is called "SelfCalibration."
The adjustment content of SelfCalibration differs depending on the color mode that is executed.
- Standard Mode (color modes excluding "CAL"):
the monitor's color reproduction gamut is updated and the display mode in each mode is adjusted as follows. - The temperature is adjusted so that it is as close as possible to the standard value.
- The color gamut values are adjusted so that they are close to each standard value.
- Information on brightness is updated.
- Advanced Mode (color mode "CAL"):
- When performing SelfCalibration on the monitor itself, calibrate the monitor to match the set targets.
- When using ColorNavigator 7, use a measurement device with the software to maintain the calibrated state of the monitor.
Attention
- Touching the internal calibration sensor can affect its measurement accuracy.
- Do not touch the sensor during use.
- Large changes in the amount of ambient light entering the internal calibration sensor light receiver can affect measurement accuracy.
- Calibration is best done in a darkened room.
- Do not put your face or objects close to the monitor, or look into the sensor, during measurement.
- Installation of a monitor hood is recommended.
- High temperatures or humidity levels can affect the measurement accuracy of the internal calibration sensor.
The monitor is best used in the following environment. - Temperature: 30˚C or below
- Humidity: 70% or below
- Avoid using in direct sunlight.
Note
- SelfCalibration can be executed 30 minutes after turning on the monitor.
- SelfCalibration can also be executed when there are no external device signals being input.
- To preserve adjustments that you set by using ColorNavigator 7, it is recommended to execute SelfCalibration.
- Because brightness and chromaticity will change as you use the monitor, it is recommended to calibrate the monitor periodically.
- The measurement results from the built-in calibration sensor can be correlated with the measurement results from the external measurement device that you want to use as a reference. For details, refer to the ColorNavigator 7 User's Manual.
You can set the calibration targets and execution schedule by using ColorNavigator 7 or the monitor's Setting menu.
You can download the ColorNavigator 7 software and User's Manual from our web site:
(www.eizoglobal.com )
Note
- When using the software, you will need to connect the external device to the monitor with the supplied USB cable.
- While using the software, do not operate the power button or control buttons on the front side of the monitor.
SelfCalibration Functions
Set the details for SelfCalibration.

Execute
SelfCalibration can be executed manually regardless of the schedule.
Note
- After selecting "Execute," warming up (after the power to the monitor is switched on, leaving the monitor on for a certain period of time until the display becomes stable) may be executed before the built-in calibration sensor comes out.
Settings - Mode Settings - Standard Mode
Settings: "On" / "Off"
Switch between enabling/disabling SelfCalibration in Standard Mode.
Settings - Mode Settings - Advanced Mode (CAL)
Settings: "On" / "Off"
Switch between enabling/disabling SelfCalibration in Advanced Mode.
Settings - Schedule - Start time
Settings: "Power Save" / "Immediately" / "Application" / "Off"
Select the timing to execute SelfCalibration upon reaching the time set in the schedule.
- "Power Save"
- To be executed under any of the following conditions.
- When the monitor is in "Power Save" mode or power is off at the set time.
- The monitor shifts to the power saving mode or the power is turned off when the set time in the schedule is exceeded.
- "Immediately"
- SelfCalibration is executed immediately at the set time.
- "Application"
- Executes SelfCalibration according to the timing set using ColorNavigator Network. For details on ColorNavigator Network, refer to our web site (www.eizoglobal.com ).
- "Off"
- SelfCalibration is not executed.
Settings - Schedule - Frequency
Settings: "Daily" / "Weekly" / "Monthly" / "Quarterly" / "Biannually" / "Annually" / "Usage Time"
Select the SelfCalibration execution cycle.
Settings - Schedule - Timing
Settings: "Jan/Apr/Jul/Oct" / "Feb/May/Aug/Nov" / "Mar/Jun/Sep/Dec" / "Jan/Jul" / "Feb/Aug" / "Mar/Sep" / "Apr/Oct" / "May/Nov" / "Jun/Dec" / "January" - "December" / "50 h" – "500 h"
When the execution cycle is "Quarterly," "Biannually," "Annually," or "Usage Time," select the timing for executing SelfCalibration.
The settings differ depending on the execution cycle setting.
- "Quarterly":
- "Jan/Apr/Jul/Oct," "Feb/May/Aug/Nov," "Mar/Jun/Sep/Dec"
- "Biannually":
- "Jan/Jul," "Feb/Aug," "Mar/Sep," "Apr/Oct," "May/Nov," "Jun/Dec"
- "Annually":
- "January" – "December"
- "Usage Time":
- "50 h" – "500 h"
Settings - Schedule - Week
Settings: "1st week" – "5th week"
When the execution cycle is "Monthly," "Quarterly," "Biannually," or "Annually," select the week for executing SelfCalibration.
Note
- When no day has been selected for "Day" in the selected week, the week of execution will be as follows:
- When the selected week is "1st week": 2nd week
- When the selected week is "5th week": 4th week
Settings - Schedule - Day
Settings: "Sunday" – "Saturday"
When the execution cycle is "Weekly," "Monthly," "Quarterly," "Biannually," or "Annually," select the day for executing SelfCalibration.
Settings - Schedule - Time
Settings: "0:00" – "23:55"
When the execution cycle is "Daily," "Weekly," "Monthly," "Quarterly," "Biannually," or "Annually," select the time for executing SelfCalibration.
Settings - Clock Adjustment
Set the monitor date and time.
Note
- When the clock has not been set, the schedule will not be applied.
- If the main power supply is disconnected for an extended period of time, the clock may need resetting.
- When ColorNavigator 7 is started, the date and time are automatically set.
Result
The SelfCalibration execution result is displayed.
Next Calibration
The next SelfCalibration scheduled execution date is displayed.
Setting Targets
Set calibration targets for Advanced Mode. You can set targets by using ColorNavigator 7 or on the monitor itself.
When using the monitor itself, set the following functions in the "Color" menu.
- "Color Mode"
- Select "CAL."
- "Target Settings"
- Set a calibration target for SelfCalibration.
Executing
SelfCalibration can be executed manually, which does it immediately, or by setting a schedule to do it automatically.
You can set a schedule by using ColorNavigator 7 or on the monitor itself.
When executing SelfCalibration by setting a schedule on the monitor itself, the method to enable SelfCalibration differs depending on the type of color mode (Standard Mode or Advanced Mode).
Attention
- If there is a change in the video signal from the external device during SelfCalibration (the signal disappears, a signal is input during a no-signal state, or other), SelfCalibration is automatically canceled.
Immediately executing manually
Manually execute SelfCalibration.
There are two methods for executing SelfCalibration, as follows.
- Executing from the "SelfCalibration" menu
- In the "SelfCalibration" menu, select "Execute."
- SelfCalibration is executed for all the color modes selected as "On" in "Settings" - "Mode Settings" of the "SelfCalibration" menu.
- Executing from the "Color" menu
- In "Color Mode" in the "Color" menu, select "CAL," and then select "Execute Calibration."
- Executes SelfCalibration for the displayed color mode.
Automatically executing by setting a schedule
Set the calibration schedule for SelfCalibration.
In "SelfCalibration" in the "Settings" menu, set the calibration schedule and monitor date and time for SelfCalibration.
Attention
- If a scheduled SelfCalibration is canceled, it is executed again when the monitor transitions to the power saving mode an hour or more after cancellation or when the monitor is turned off using
.
Activating SelfCalibration in Standard Mode
Execute SelfCalibration in Standard Mode.
In "Settings" in the "SelfCalibration" menu, set the following functions:
- "Mode Settings"
- Set "Standard Mode" to "On."
Activating SelfCalibration in Advanced Mode
Execute SelfCalibration in Advanced Mode.
In "Settings" in the "SelfCalibration" menu, set the following functions:
- "Mode Settings"
- Select "Advanced Mode."
- "Advanced Mode"
- Select "CAL," and set "On."
Checking Results
Check the SelfCalibration calibration results by using the monitor.
In the "Color" menu, set the following functions:
- "Color Mode"
- Select "CAL."
- "Result"
- Check the calibration result of SelfCalibration.