Monitor

Compatibility between RadiForce monitors using DisplayPort daisy chains

We confirm that the following combinations of DisplayPort daisy chains are possible.

- Multi monitor connection diagram

Multi monitor connection diagram

 

- 4-screen monitor configuration (Some graphics boards may not support this configuration)

4-screen monitor configuration

 

 

- Supported monitors

RadiForce RX660 RadiForce GX560, GX550, RX560 RadiForce RX360, RX350 RadiForce RX250, MX216    
RadiForce MX315W          
 

RadiForce RX660

 

Dual monitor configuration - Pattern 1

Dual monitor configuration

  • Set [Preferences] - [Input] to 'DisplayPort 1'.
  • Set [Administrator Settings] - [Daisy Chain] to 'On'.
  • This pattern does not support 10-bit.
Graphics board Monitor 1 Monitor 2
MED-X90
MED-X70
MED-X50LP
MED-X30LPB
MED-XN91
MED-XN71
MED-XN51LP
MED-XN31LP
RadiForce RX660 RadiForce MX216
RadiForce MX215
FlexScan EV2456
FlexScan EV2455
FlexScan EV2451
FlexScan EV2450
FlexScan S2133
FlexScan S1934


 

 

Dual monitor configuration - Pattern 2

Dual monitor configuration

  • Set [Preferences] - [Input] to 'DisplayPort 1'.
  • Set [Administrator Settings] - [Daisy Chain] to 'On'.
  • Set [Preferences] - [PinP] - [PinP Display] to 'On' and set [Preference] - [PinP] - [Input] to 'DisplayPort 2'.
  • This pattern does not support 10-bit.
Graphics board Monitor 1 Monitor 2
MED-X90
MED-X70
MED-X50LP
MED-XN91
MED-XN71
MED-XN51LP
MED-XN31LP
RadiForce RX660 Connect the monitor directly to the graphics board, not via a daisy chain.

Page Top


 

RadiForce GX560, GX550, RX560

 

3-screen monitor configuration


3-screen monitor configuration
 

Graphics board Monitor 1 Monitor 2 Monitor 3
MED-X90
MED-X70
MED-X50LP
MED-XN91
MED-XN71
MED-XN51LP
MED-XN31LP
RadiForce GX560
RadiForce GX550
RadiForce RX560
Connect the monitor directly to the graphics board, not via a daisy chain.


Page Top


 

RadiForce RX360, RX350

 

3-screen monitor configuration - Pattern 1

 

3-screen monitor configuration
 

Graphics board Monitor 1 Monitor 2 Monitor 3
MED-X90
MED-X70
MED-X50LP
MED-XN91
MED-XN71
MED-XN51LP
MED-XN31LP
RadiForce RX360
RadiForce RX350
RadiForce MX216*
RadiForce MX194*
FlexScan EV2456
FlexScan EV2455
FlexScan EV2451
FlexScan EV2450
FlexScan S2133
  • If Monitor 3 is MX216, set [Administrator Settings] - [DisplayPort] to 'Version 1.1 8bit'.
  • Only if Monitor 1 and 2 are RadiForce RX360
     
 

4-screen monitor configuration - Pattern 2

 

4-screen monitor configuration
 

Graphics board Monitor 1 Monitor 2 Monitor 3 Monitor 4
MED-XN91
MED-XN71
MED-XN51LP
MED-XN31LP
RadiForce RX360
RadiForce RX350
RadiForce MX216*
RadiForce MX194*
FlexScan EV2456
FlexScan EV2455
FlexScan EV2451
FlexScan EV2450
FlexScan S2133
Connect the monitor directly to the graphics board, not via a daisy chain.
  • If Monitor 3 is MX216, set [Administrator Settings] - [DisplayPort] to 'Version 1.1 8bit'.
  • Only if Monitor 1 and 2 are RadiForce RX360


Page Top

 

RadiForce RX250, MX216

 

3-screen monitor configuration - Pattern 1

3-screen monitor configuration

 

Graphics board Monitor 1 Monitor 2 Monitor 3
MED-X90
MED-X70
MED-X50LP
MED-XN91
MED-XN71
MED-XN51LP
MED-XN31LP
RadiForce RX250
RadiForce MX216
RadiForce MX242W
RadiForce MX216
RadiForce MX215
FlexScan EV2456
FlexScan EV2455
FlexScan EV2451
FlexScan EV2450
FlexScan S2133

 

 

 

4-screen monitor configuration - Pattern 2

4-screen monitor configuration

 

Graphics board Monitor 1 Monitor 2 Monitor 3 Monitor 4
MED-XN91
MED-XN71
MED-XN51LP
MED-XN31LP
RadiForce RX250
RadiForce MX216
RadiForce MX242W
RadiForce MX216
RadiForce MX215
FlexScan EV2456
FlexScan EV2455
FlexScan EV2451
FlexScan EV2450
FlexScan S2133
Connect the monitor directly to the graphics board, not via a daisy chain.


Page Top

 

RadiForce MX315W

 

Dual monitor configuration - Pattern 1

Dual monitor configuration

  • Set [Preferences] - [Input] to 'DisplayPort 1'.
  • Set [Administrator Settings] - [Daisy Chain] to 'On'.
  • This pattern does not support 10-bit.
Graphics board Monitor 1 Monitor 2
MED-X90
MED-X70
MED-X50LP
MED-X30LPB
MED-XN91
MED-XN71
MED-XN51LP
MED-XN31LP
RadiForce MX315W FlexScan S1934

 

 

Dual monitor configuration - Pattern 2

Dual monitor configuration

  • Set [Preferences] - [Input] to 'DisplayPort 1 l DisplayPort 2'.
  • Set [Administrator Settings] - [Daisy Chain] to 'On'.
Graphics board Monitor 1 Monitor 2
MED-XN91
MED-XN71
MED-XN51LP
MED-XN31LP
RadiForce MX315W Connect the monitor directly to the graphics board, not via a daisy chain.


Page Top